Home > Product Center > Power Electronic Tester > Programmable DC power supply > High-precision Programmable Linear DC Power Supply
TH6502 High-precision Programmable Linear DC Power Supply

RS232、USBHOST、USBDEVICE

GPIB

Performance characteristics

 Output: 32V/3A(96W)

■ High precision and resolution: 0.1mV/0.01mA

■ High speed: 1ms

■ Voltage test, resistance test (build-in 5 1/2 digital 

   milliohmeter)

■ LCD resolution 480*272

■ Numeric keypad operation

 Low ripple and low noise

 Intelligent fan control to save energy and reduce noise

 Software monitoring via computer

 Editable voltage and current output waveform with time

   (resolution 1ms) (LBT mode)

 The power output can be turned on and off by an external

   signal 

 The knob can be used to coarsely adjust and fine tune the 

   voltage and current values. 

 Timing output time can be set (0.01-9999.99S)

 Screen information can be stored in the USB flash drive

 Chinese and English user interface

 Flexible and convenient file operating system

 Automatic upgrade of instrument operating software via USB 

   HOST

 Handler interface for online operations 

 RS232, USB HOST, USB Device, GPIB can easily realize 

   the data communication with PC and remote control of the  

   instrument

 Comes with hardware OVP, OCP protection

   (OCP is software protection)

Brief Introduction
Application

 R & D and design verification common test

 Production line table routine testing and maintenance

 Automated device integration testing

 Solar photovoltaic simulation test

 New power car simulation test 

 Teaching laboratory



Technical Parameter
ModelTH6501TH6502TH6503TH6511TH6512TH6513
Rated OutputVoltage0-20V0-32V0-72V0-20V0-32V0-72V
Current0-5A0-3A0-1.5A0-10A0-6A0-3A
Power100W96W108W200W192W216W
Load RegulationVoltage≤0.01%+2mV
Current≤0.05%+1.5mA
Power RegulationVoltage≤0.01%+1mV
Current≤0.05%+1mA
Set Value ResolutionVoltage1mV
Current0.1mA
Read-back ResolutionVoltage0.1mV
Current0.01mA
Year Set Accuracy (25℃±5℃)Voltage≤0.03%+3mV
Current≤0.05%+2mA
Year Read-back Accuracy(25℃±5℃)Voltage≤0.02%+3mV
Current≤0.05%+2mA≤0.05%+2.5mA
Ripple and Noise(20Hz-20MHz)Differential Mode Voltage≤3mVp-p and 1mVrms≤4mVp-p and 1mVrms
Differential Mode Current<3mArms<4mArms
Dynamic Recovery Time (50%-100% LOAD)Restore Time within 75mv<200us
Rise Time10%-90%<20ms
Fall Time90%-10%<200ms<250ms<150ms<200ms<250ms<150ms
Overvoltage ProtectionRange (Typical)1-19V1-31V1-71V1-19V1-31V1-71V
Accuracy (typical)± (set value *0.5%+0.5V)
Response Time (typical)<10ms
DVM(DC)Display Value Accuracy±0.02%+10mv
Display Resolution0.1mv
Input Differential Mode Voltage Range0-40Vpk
Input Common Mode Voltage Range0-30Vpk


Accessary
Standard
Accessories nameModel
Test line at both ends TH26004C
Optional
Accessories nameModel
High current test cable TH26035D
High current test cable TH26035E
Data download
Name of dataDownload Link
SpecificationDownload
TH6500 Series Operation Manual Download
Return to the top