TH1952 Digit Multimeter
Performance characteristics

120000 count display

Max. ADC sampling rate can be 200 times/s

 High brightness VFD dual-display

 Standard Handler interface, provide external trigger and 

   sorting signal for production line

 True-rms AC voltage and current test

 Multiple mathematics operation function DCV/ACV/DCI/

   ACI/AC+DC/Ω/CAP/Temperature/Frequency/Diode/

   Continuity and dB/dBm

 Other parameters in current test signal displayed in 

   second display

 DCV test accuracy up to 0.012%,resolution up to1μV

 Relative mode(REL)to eliminate remaining reading

 Calibration without opening the case

 Limit function(Limit), for fast sorting

 Data statistics function Max/Min, Max. value, Min. value,

   Average value and number of statistics

 Standard USB interface,provide convenient system 

   communication function

Brief Introduction

■ TH1952 5 1/2 digit multimeter is a multi-functional, costeffective voltage/current/resistance/capacitance meter. It provides Max. 120000 count, Max. sampling rate 200 times/s, 0.012% DCV accuracy ,etc, which is the optimal selection for users.

TH1952 provides high brightness VFD Dual-display, which can display ACV/DCV or DCI, ACV/ACI and frequency .etc to improve the test efficiency and, clearer display.

TH1952 is configured with USB interface, and also communication software is provided for PC communication, data analysis, statistics, building system. TH1952 accepts SCPI command to be compatible with communication software. Also Handler interface is provided to receive the external trigger and output sorting signal for production line.


Application
Technical Parameter
Test  Function
Test ParameterDCV, ACV, DCI, ACI, DCR, CP, FREQ, CONT, DIODE, TEMP
Mathematics Function%, dB, dBm, REL, mX+b
RangeAuto, Manual
DisplayHigh brightness VFD dual-display
Trigger ModeINT/BUS/EXIT
Reading HoldTo find out best stable reading for each data block of the given reading number according to the given accuracy.
LimitationMeasurementTo judge HI、IN、LO and display, with ALARM for HI/LO
CalibrationAuto, Manual
External TriggerHandle interface (external trigger and output sorting signal)
Communication InterfaceSCPI command support for RS232 and  USB interface
Performance parameter  expression of uncertainty: ±(% of reading+% of full scale)
Reading Rate(reading/s)
Reading RateFastSlow
1/21/21/21/2
DCV/DCI8015154
ACV/ACI8015154
Ω8015154
DCV
RangeMin. ResolutionMax. Uncertainty(1 year)
100mV-1000V1µV0.012%+0.008%
DCI
RangeMin. ResolutionMax. Uncertainty(1 year)
1mA-100mA10nA0.05%+0.005%
1A10µA0.10%+0.008%
10A100µA0.20%+0.006%
ACV
RangeFrequencyMin. ResolutionMax. Uncertainty(1 year)
100mV-750V20Hz-50Hz1µV1.00%+0.1%
50Hz-10kHz0.20%+0.08%
10kHz-30kHz0.80%+0.08%
30kHz-100kHz3.00%+0.2%
ACI
RangeFrequencyMin. ResolutionMax. Uncertainty(1year)
1mA -10A20Hz-50Hz10nA1.0%+0.08%
50Hz-2kHz0.5%+0.08%
2kHz-10kHz2.0%+0.18%
Resistance
RangeMin. ResolutionMax. Uncertainty(1 year)
100Ω-1MΩ10µΩ0.05%+0.008%
10MΩ100Ω0.20%+0.008%
100MΩ1kΩ2.00%+0.005%
Capacitance
RangeMin. ResolutionMax. Uncertainty(1 year)
1nF1pF2%+0.8%
10nF-1000uF10pF1.0%+0.5%
10000uF1uF2%+0.5%
Frequency
RangeVoltage sensitivityMin. ResolutionMax. Uncertainty(1year)
1Hz-1MHz40mV rms10µHz0.005%+0.003%
Temperature
RangeProbe typeMin. ResolutionMax. Uncertainty(1year)
-10℃-100℃PT5000.1℃Probe accuracy±0.2℃
Uncertainty is not assured if the temperature is out of the range


Accessary
Standard
Accessories nameModel
Multimeter test cable TH26036(JC-BBA03601)
Optional
Accessories nameModel
TH1951/61 four-terminal test line TH26039
Data download
Name of dataDownload Link
SpecificationDownload
TH1952 Operation Manual Download