Performance characteristics

 Freq: 20Hz-300kHz, resolution: 1mHz max.  

 Test Accuracy: 0.05%

 Test Speed: 9ms

 800×RGB×480  7-inch TFT LCD display

 Strongest signal source selection:

   10V/100mA programmable AC test level

   10V/100mA programmable DC bias supply

   10V/50mA standalone DC voltage source

   1A interior DC bias current source (optional)

   120A external bias source (optional)

 Simultaneous display of 4 kinds of test parameters

 201 -point list sweep function

 Continuous curve scanning/graphical analysis function 

 Internal storage of 100 sets of LCRZ setting files and 10 

   sets of GIF image

 GIF image and CSV data files can be saved to USB storage 

   directly

 HANDLER, USB, LAN, RS232C, GPIB (option), DCI 

   interface


Brief Introduction

By dint of leading impedance measurement technology and rich R&D experience, Tonghui continuously introduces representative impedance measurement product --- TH2829 series automatic component analyzer is another excellent product we have produced. TH2829A automatic component analyzer possesses a higher test speed, a more comprehensive analysis function and friendly human-computer interactive experience by adopting the latest high-speed processor and a new software system. Well-designed measuring circuit and optimized algorithms further enhance the test stability of low-D capacitance and high-Q inductors. The instrument is provided with 10V AC test level, 10V/100mA bias current and standalone 10V/50mA DC current, making it convenient for applying in the test of all kinds of active/ passive devices. Main/ sub parameters display, enhanced display system design, 150-points list sweep and graphical analysis capabilities of multiple parameters meet the most application requirements of customers.

Thanks to the application of a new generation of processors, the instrument has a more powerful data processing capability. The test results can be easily stored in the U disk or uploaded to the upper PC or network through multiple interfaces, promoting test automation and test efficiency.

The test frequency of TH2829A is 20Hz-300kHz.  The instrument has a test accuracy of 0.05% and highest test speed of 9ms/time. Being equipped with multiple interfaces of HANDLER, USB, LAN, RS232C, DCI, GPIB (option) as well as rich resources, the instrument will bring excellent cost performance experience for customers.

TH2829 series automatic component analyzer is completely appropriate for test requirements of all kinds of industrial and military standards.



Application
Technical Parameter
ModelTH2829A
Frequency of test signal
20Hz—1MHz
Minimum resolution1mHz, 5-digit frequency input
Accuracy0.01%
Display800×RGB×480  7-inch TFT LCD display
ACLevelVoltage range of test signal5mV—10Vrms
Minimumresolution of voltage100μV, 3-digit input
AccuracyALC ON10% x set voltage + 2mV
ALCOFF6% x set voltage + 2mV
Current range of test signal50μA—100mA
Minimumresolution of current1μA, 3-digit input
AccuracyALC ON10% x set current + 20μA
ALCOFF6% x set voltage + 20μA
DCbias voltage sourceVoltage/Current range0V— ±10V / 0mA—±100mA
Resolution0.5mV / 5μA
Voltage accuracy1% x set voltage + 5mV
ISO ONBe used for the bias test of inductance and transformer
AC Source impedanceISO ON100Ω
ISOOFF30Ω、50Ω、100Ωselectable
DCR Source impedance30Ω、50Ω、100Ωselectable
DCIndependent voltage sourceVoltage/current range0V— ±10V / 0mA—±50mA
Resolution0.5mV / 5μA
Voltage accuracy1% x set voltage + 5mV
Output resistance100Ω
Test parameters of LCR|Z|,Y|, C, L, X, B, R, G, D, Q, θ, DCR,Vdc-Idc
Parameter display of test pageTwo sets of main/sub parameters, the second set can be set as ON/OFF;There can be 10 pages of list sweep and 15 points per page at most;Multiple parameters continuous sweep graphical analysis.
Basic accuracyLCR test parameter0.05%
CalibrationWarm-up time ≥ 30 seconds;Environment temperature: 23±5ºC;  Signal voltage: 0.3Vrms-1VrmsZeroing: After OPEN or SHORT;    Length of test cable: 0 m
Measurement time(≥10 kHz)Fast: 9 ms / timeMedium: 67 ms / timeSlow:187 ms / time Plus the refresh time of display character
Display range ofLCR parameterZ, R ,X,DCR0.00001Ω — 99.9999MΩ
|Y|,G,B0.00001μs — 99.9999s
C0.00001pF —  9.99999F
L0.00001μH —  99.9999kH
D0.00001 —  9.99999
Q0.00001 —  99999.9
θ(DEG)-179.999º — 179.999º
θ(RAD)-3.14159 — 3.14159
Δ%-999.999% — 999.999%
Equivalent circuitSerial, Parallel
Range modeAuto, Hold
Trigger modeInternal, Manual, External, Bus
Average times1-256
Calibration functionOpen, short calibration with full frequency or dot frequency, Load
Math operationDirect reading, ΔABS, Δ%
Delay time setup0-999, minimum resolution: 100us
Comparator10-bin sorting, BIN1-BIN9, NG, AUX
Bin counter
PASS/FAIL on front panel, LED indication
List sweep·201 -point list sweep function ·List sweep of frequency, AC voltage/  current, internal/external DC bias  voltage/current and independent DC  source voltage can be performed on  each page. Each sweep point can be  sorted separately.
Graphical analysis·Graph scanning and analysis of  frequency, AC level and DC bias can  be performed.·Set the sweep start point, end point  and each sweep point.·Display the maximum value, minimum  value and read any of the chosen   sweep point·Scanning graphs can be stored into  internal or external USB memory.
Internal nonvolatilememory100 sets of LCRZ setting files memory201 times test results10 sets of GIF image, CSV data files
External USB memory·GIF image, CSV data files·LCRZ setting files memory ·Test data can be stored via USB   memory directly.
Interface1A bias current source1A DC bias current source (optional) can be stalled
I/O interfaceHANDLER on rear panel
SCIUSB、RS232C
PCIGPIB(optional)
NILAN
Memory interfaceUSB HOST(front panel)
General Specifications
Operating temperature and humidity0℃-40℃, ≤ 90%RH
Power supplyVoltage99V-121V, 198V-242V AC
Frequency47Hz-63Hz
ConsumptionMax. 80 VA
Dimension(W×H×D)400mm × 132mm × 385mm
WeightApprox.13 kg


Accessary
Standard
Accessories nameModel
Fixture TH26048
Short circuit TH26010
Kelvin test lead with box four-terminal insulation and lock TH26011AS