■ Number of Test PIN: 20
■ Frequency: 20Hz-200kHz, Resolution: 0.5mHz
■ Signal Level: 5mV-2Vrms(Option:2mV-10Vrms)
■ Test Speed: max. 13ms
■ 7-inch TFT LCD display with a resolution of 800×RGB×480
■ Built-in 0-100mA/0-10V bias power supply, optional 1A/2A
bias current source
■ Diode forward and reverse characteristic detection
■ Improved high turns ratio and weakly coupled transformer
test capability
■ Improved DCR testing capabilities
■ Single screen can accommodate all scan test results
■ Sort the selected scanning parameters
■ Time stamping system: memory file setting, calibration
deviation and deduction time
■ Self-test scanning fixture relays
■ Flexible deviation deduction method
■ Multiple handling ways for FAIL cases
■ Single parameter test cycle to test independent windings
■ Increased security: administrator and operator passwords
■ Built-in statistical analysis capabilities: Cpk, Cp, Ck, etc.
■ Bar-code reading function can be used to select a setting file
or to manage the type of test products
■ Optional PC-level instrument test setup file programming
capability
■ Online upgrade mode: USBHOST or RS232
■ Support multiple instrument networking through LAN
interface
■ Backward compatible with TH2818X/TH2819X parameter
setting file
■ Storage: Internal: 100 groups of settings file to save
U disk: 500 groups of configuration files, CSV
format test data, GIF format images
■ Switching transformer scanning test, comprehensive characteristics analysis.
■ Network transformer scanning test, comprehensive characteristics analysis
■ Discrete passive components (L, R, C) multi-channel scanning test
■ Relay drive line package, contact resistance multi-channel scanning test
■ Multi-channel DC resistance DCR scanning test
■ Comprehensive test analysis of multiple passive components in impedance network
Model | TH2829LX | TH2829AX | TH2829AX-24 | TH2829AX-48 | TH2829NX | TH2829CX |
Test Pin(PIN) | 20 | 20 | 24 | 48 | 72/96/120/144/168/192 | 20 |
Test Frequency | 20Hz — 200kHz | 20Hz — 1MHz | ||||
Display | 800×RGB×480 7 inch TFT LCD display | |||||
LCR Function | option | |||||
Transformer test parameters | Turn Ratio, Turns, Phase, L, C, Lk, Q, ACR, DCR, Balance, Pin Short, Diode P/N | |||||
LCR Test Parameters | |Z|,Y|, C, L, X, B, R, G, D, Q, θ, DCR, Turn-Ratio, Phase, Lk | |||||
Basic Test Accuracy | LCRZ | 0.05% | ||||
DCR、Turn Ratio | 0.1% | |||||
Signal Source Output Impedance | 10Ω、30Ω、50Ω、100Ω | |||||
Test Speed (ms/times) | 13ms, 90 ms, 370 ms | |||||
AC Signal Level | 5mVrms — 2Vrms(transformer test, can be customized to 10Vrms), 5mVrms — 10Vrms(LCR function);50μArms — 100mArms | |||||
DC Bias Voltage Source | ------ | 0V — ± 10V;0mA — ± 100mA | ||||
DC Bias Current Source | 0 — ±1A option(option TH2901) / 0 — ±2A option(option TH2902) | |||||
DC Constant Current Source | 0mA – ±120mA for diode forward characteristic test | |||||
Diode Test | forward test voltage | 0 — 9.9999 V | ||||
Reverse test current | 0 — 99.999 mA | |||||
Comparator | 10 bins, PASS/FAIL indication, file counting function | |||||
Storage | Internal: 100 sets of configuration file;U disk: 500 sets of configuration files, CSV format test data, GIF format images |
Optional | |||||
Accessories name | Model | ||||
Pneumatic transformer scanning test fixture | TH1901A | ||||
Manual transformer scan test fixture | TH1901C |