■ Test Voltage: Continuously adjustable 1-800V,
and real-time monitor the output voltage
■ Voltage Accuracy: ±(0.5% set value+0.2V)
■ Test Range: LC: 0.001μA-20.00mA
IR: 0.01MΩ-99.99GΩ
■ Current Accuracy: ±(0.3%+0.05μA)
■ Test Speed: Max. 40ms
■ Capacitance leakage test function
■ Insulation resistance test function
■ Aluminum foil pressure and rise time test function
■ Precise low current charge function(0.5mA±0.05mA)
■ Large current (500mA) improves the charge speed of low
voltage large capacitance.
■ Open correction(null) to eliminate the remaining base
number
■ Built-in digit counter
■ Comparator function to realize the sorting of PASS/FAIL
■ 10 groups of status for save and load
■ Standard RS232 interface, Handler interface, optional GPIB
interface
■ Large LCD ( 240×64 dot-matrix )display
■ Humanized operation interface
■ TH2689/TH2689A is a rapid and precise capacitance leakage current insulation resistance tester. The max. test speed of 18 times/second, good performance and the suitable price provide the user an optimal choice.
TH 2689/TH2689A provides max. test voltage: 800v/500v, charge current: 0.5mA─500.0mA( if >100V, the max. power 50W can limit). It is mainly applied in capacitance leakage current, insulation resistance and aluminum electrolysis capacitance anode foil pressure test. Also it can be applied in the confirmation of annihilator, zener diode, neonbulb .etc and leakage current test. Standard Handler interface, stable and rapid test, to reach the sorting effect
Standard RS 232 interface, optional GPIB interface, and general RS232 software are provided to meet the demand of connecting with computer and forming auto test system. The instrument can receive the SCPI command to guarantee the compatibility of communication software.
Model | TH2689 | TH2689A |
Function: LC/IR test | ||
Test voltage | 1.0V ─ 800V | 1.0V ─ 500V |
Accuracy: ±(0.5% set value+0.2V) | ||
Charge current | test voltage ≤ 100V,0.5mA─500mA; test voltage > 100V, 0.5mA─Imax, Imax=50W/test voltage | |
Accuracy: ±(3% set value+0.05mA) | ||
Test range | LC: 0.001uA ─ 20.00mAIR: 0.01kΩ ─ 99.99GΩ | |
Basic accuracy | LC: ±(0.3%+0.05uA) | |
Charge time upper limit | 0 ─ 999s manual | |
Test time | FAST: 40ms; MED: 60ms; SLOW: 120msTest condition: range is locked, trigger mode is EXT and the external trigger voltage displays the closing state | |
Limit setting | LC: 0 ─ 999.999mAIR: 0 ─ 999.999GΩ | |
Sorting | Pass, Fail | |
Function: W.V. test | ||
VfRated involucra voltage | TH2689: 1.0V ─ 800V | TH2689A: 1.0V ─ 500V |
Charge current | 0.5mA ─ ImaxImax = 65W/VfAccuracy: ± (3% set value+0.05mA) | |
Charge Time Upper Limit | 5s ─ 600s manual | |
Pressure Time | 30s ─ 600s manual |
Standard | |||||
Accessories name | Model | ||||
Three-terminal test line | TH26004D |
Optional | |||||
Accessories name | Model |