■ Freq: 20Hz-5MHz, max. resolution 10mHz
■ Test Accuracy: 0.1%
■ Test Speed: 200 times/s
■ Multi-parameter with multi-frequencylist sweep function
■ Test level:10mV-5V with the resolution of 1mV
■ 320×240 dot-matrix large graphic LCD display
■ 5-digit display resolution
■ 22 parameter combinations available
■ 4 signal source output impedance
■ 10 points list sweep function
■ Built-in DC bias source
■ External AC bias up to 40A (option)
■ Auto level control (ALC) function of voltage or current
■ V,I test signal level monitor function
■ Graphic scanning and analyzing function
■ 20 groups of setting for storage/load
■ Built-in comparator, 10-bins and bin counters
■ Multiple communication interfaces
■ 2m/4m cable length extension
■ Optional Chinese and English language operating interface
■ Software upgrade through USB HOST
TH2826 series is a new generation impedance test instrument and firstly in conformity with LXI standard inland. It fulfills most low-voltage parameter measurement needs for components and materials with its basic accuracy (0.1%), wide frequency range(from 20Hz to 5MHz). The instrument is widely applied in microphone, resonator, inductor, ceramic capacitor, LCD, varactor and transformer .etc to analyze electrical performance and test low-ESR capacitor and high-Q inductor. With the super high test speed, TH2826 series is especially for frequency response curve analyzing of detector and piezoelectric device on auto production line. Its multiple output impedance mode can be used for different standards of inductor or transformer manufacturers.
With its prominent performance, TH2826 series is in conformity with commercial and military standards, such as IEC and MIL standards.
Model | TH2826 | |||
Test frequency | 20Hz–5MHz with resolution of 10mHz | |||
Measurement Parameters | C, L, R,Z,Y,X,B, G, D, Q, θ, DCR | |||
Test Level | f≤1MHz | 10mV–5V,±(10%+10mV) | ||
f>1MHz | 10mV–1V,±(20%+10mV) | |||
Output impedance | 10Ω, 30Ω, 50Ω, 100Ω | |||
Basic Accuracy | 0.1% | |||
Display Range | L | 0.0001 uH – 9.9999kH | ||
C | 0.0001 pF – 9.9999F | |||
R,X,Z,DCR | 0.0001 Ω – 99.999 MΩ | |||
Y, B, G | 0.0001 nS – 99.999 S | |||
D | 0.0001 – 9.9999 | |||
Q | 0.0001 – 99999 | |||
θ | -179.99° – 179.99° | |||
Measuring Speed(meas/sec) | Fast: 200(f>30kHz),100(f>1kHz)Med: 25, Slow: 5 | |||
Calibration function | Open/Shot /load | |||
Equivalentmode | Serial, Parallel | |||
Ranging Mode | Auto and Hold | |||
Display Mode | Direct, ABS, Rel | |||
Trigger Mode | Internal, Manual, External, BUS | |||
Internal DC bias source | Voltagemode | -5V – +5V,±(10%+10mV), with the resolution of 1mV | ||
Currentmode(internal resistance is50Ω) | -100mA – +100mA, ±(10%+0.2mA), with the resolution of 20uA | |||
Comparator function | bins and bin counters | |||
Display | 320×240 dot-matrix LCD display | |||
Memory | groups of control settings can be saved | |||
Interface | USB DEVICE( USBTMC and USBCDC support) USB HOST(FAT16 and FAT32 support) LAN(LXI class C support) RS232C HANDLER GPIB(option) | |||
Working Temperature & Humidity | 0°C − 40°C, ≤90%RH | |
Power supply | Voltage | 99V−121V,198V−242V |
Frequency | 47.5Hz−63Hz | |
Power Consumption | ≤ 80VA | |
Dimensions (W×H×D) | 400mmx132mmx385mm | |
Weight | Approx. 10kg |
Standard | |||||
Accessories name | Model | ||||
Fixture | TH26048 | ||||
Short circuit | TH26010 | ||||
Kelvin test lead with box four-terminal insulation and lock | TH26011BS |
Optional | |||||
Accessories name | Model |