TH2684 High Precision IR Tester

RS232、USBHOST、USBDEVICE、HANDLER

GPIB

Performance characteristics

■ IR resistance test range: 10kΩ - 50TΩ

■ Charge current: 2mA , 25mA, 200mA 

■ Test speed: 50ms

■ Accuracy: ±2% (< 1TΩ ) 

■ Test voltage range: 10-500V, 1V step

■ 320×240 dot-matrix LCD

■ Ultra-low leakage current test: minimum current is 

   10pA, accuracy: 2% ±2pA

■ Dual outputs (precharge voltage output and test 

   voltage output) can be set.

■ The precharge voltage output can be set to follow the test 

   voltage output and can be finely adjusted on test voltage. 

   Also the precharge voltage can be set to work in

   independent mode.

100W powerful charging function

Contact detection function for capacitive components

When the test current is less than 10nA, the internal input

   impedance can be selected between 10kΩ and 1MΩ to 

   ensure rapid and accurate test.

7 current ranges, manual or auto range mode

4-bin comparison function

Programmable sequence test mode

R-T and I-T Curve test and display mode

Screen hardcopy to be saved as BMP file to a U disk

Automatically upgrade firmware by a U disk

■ Achieve automatic test system by Handler interface

Achieve remote control by RS232C and USB 

   Device interface

Support scanning interface for mass tests

■ Auto store setup parameters

Selectable Chinese and English operation interfaces

Brief Introduction

TH2684/TH2684A High Precision IR Tester is an intelligent measurement instrument that is used for rapid measurements on IR properties of electronic parts and components, dielectric materials, equipments, cables, etc. Large LCD and user friendly menu provide you easier operation.

This instrument is especially designed for capacitor IR test TH2684/TH2684A can achieve rapid measurements through following methods:

1. Selectable internal input impedance: If the current is greater than 10nA, only 10kΩ input impedance can be used; if the current is below 10nA, you can choose 10kΩ or 1MΩ impedance to test.

2. With the built-in dual voltage output, TH2684 can charge large capacitors. By dual voltage output, TH2684 is able to output a precharge voltage up to 500V, 200mA. In voltage follow mode, precharge voltage follow with the test voltage output and can be finely adjusted. Above features ensure the perfect charge of capacitive materials.

3. TH2684A can output a voltage of 1000V, 100mA to fully charge the capacitive material. In addition, user can program the sequence measurement steps (up to 18 steps) on TH2684/TH2684A. For instance, charge, wait, test, and discharge steps can be programmed. Each step can last up to 100s. 

TH2684/TH2684A has a unique contact detection function. For capacitive material such as capacitors and cables, contact detection function can detect the contact of components under test. Moreover, this detection function will not increase any test time. 

TH2684 equips with interfaces of RS232, USB DEVICE, SCANNING and Handler. Handler interface provide convenience for automatic test system; SCANNING interface is useful for mass measurement of components. User can use a scanner to speed measurement of components.



Application
Technical Parameter
ModelTH2684TH2684A
Resistance Test
RangekΩ to 50TΩkΩ to 100TΩ
AccuracyTest current > 100pA: 2%Test current ≤ 100 pA: 2% ± Vtest/2pA
Current Test
RangeRange 1 :100uA – 1mA , Internal Input impedance 10 kΩ
Range 2 :10uA – 100uA , Internal Input impedance 10 kΩ
Range 3 :1uA – 10uA , Internal Input impedance 10 kΩ
Range 4 :100nA – 1uA , Internal Input impedance 10 kΩ
Range 5 :10nA – 100nA, Internal Input impedance 10 kΩ
Range 6 :1nA – 10nA, Internal Input impedance 10 kΩ or 1MΩ (selectable)
Range 7 :10pA – 1nA, Internal Input impedance 10 kΩ or 1MΩ (selectable)
Accuracy2% ± 2pA
Measurement Voltage
Range10 to 500V, 1V resolutionto 1000V,1V resolution
Accuracy2% of readout,or ± 1V
Source Resistance200Ω
Current Limit2,25,or 200mA2, 25 , or 100mA
Voltage OutputManually turn on or off on front panel, or controlled by built-in timer, or by remote control.
TimingProgrammable charge time: 0 to 1000s
Measurement Delay0 to 1000s programmable
Discharge Resistance2kΩ
Discharge Timet = 0.03 x Cx (in μF), when Vtest falls to 1% of the test level.
Measurement Speed
Trig ModeSingle measurement: < 100ms(exclude charge time)Average up to 100 measurements:<100+ (N-1) x 100 ms (exclude charge)
Continuous ModeDirect readout: 100ms – 10000ms depending on average number
Comparatorbins:(3 bins for PASS,1 bin for FAIL)
Range ModeAuto, Hold
Average Timesto100
MemorySets of setup values can be stored.
Standard InterfaceGPIB (optional), RS232C, HANDLER interface output, USBDEVICE ( USBTMC and USBCDC support), USBHOST, SCANNING




Operating Temperature and Humidity10°C - 40°C,  ≤90%RH
Power Supplyto 130 V  AC(60Hz) orto 260V  AC(50HZ)
Power ConsumptionTH2684 : 250WTH2684A: 150W
Dimensions (W×H×D)430mm×400mm×130mm
WeightTH2684 : 14kgTH2684A: 10kg